Mercury Monitoring System
Automated Mercury Monitoring from 2 to 24 sample points
The MMS mercury monitoring system has changed the way companies approach complete facility monitoring of mercury. The system is designed to automatically and sequentially monitor mercury from 2 to 24 sample points. A special version employing the Mercury UltraTracer UT-3000 as detector is available for concentrations in the sub-µg/m³ range. Options like air conditioning assure reliable function of the system even under extreme climatic conditions.
Benefits of the MMS Mercury Monitoring System:
Functionality of the Multi Point Mercury Monitor
The sample is continuously drawn from each sampling point through special tubing to a heated multiplexer
valve unit. One sample at a time is fed into the Mercury Vapor Monitor where the mercury concentration is measured. All sample lines are permanently purged with sample thus minimizing the response time of the system. By this method sample line lengths of 100 m and even more can be employed.
Mercury Vapor Monitoring System Design
The MMS system basically comprises a multiplexer-valve unit and a highly accurate and selective mercury detector. A bypass pump maintains a steady flow of sample through the sample filters, sample lines and multiplexer valves. Flow meters with optical alarm sensors guarantee that flow rates are stable. All components are installed inside a protective cabinet. A special version employing the Mercury UltraTracer UT-3000 as detector is available for concentrations in the sub-µg/m³ range. Options like air conditioning assure reliable function of the system even under extreme climatic conditions.
Applications for the MMS Include:
Click Brochure Above to Download PDF
Technical Specifications of the MMS
Multiplexer Module |
|
| Number of sampling points: | 2 ... 24 |
| Purging of sample lines: | continuously, ca. 40 ... 80 l/h for each channel |
| Flow control: | Flow meters with alarm sensors for low flow |
| Heating of valve unit: | approx. 70 °C |
| Sample line connectors: | fittings for 4 mm i.d./ 6 mm o.d. tubing |
| Valves: | minimum dead volume |
| Measuring duration: | 45 sec ... 99 minutes, adjustable separately for each channel |
Detector |
|
| Measuring principle: | UV absorption at 253,7 nm |
| Type for standard range: | Mercury-Vapor-Monitor VM-3000 |
| Measuring range (standard): |
0.1 to 100 µg/m³; 1 to 1000 µg/m³; 1 to 2000 µg/m³ |
| Type for low range: | Mercury Ultratracer UT-3000 |
| Measuring range (low range): | 0.01 to 50 µg/m³ |
| Sample flow: | ca. 80 l/h |
| PC connection: | serial protocol RS 232; connection by cable (optionally by fibre optic cable); modbus RTU; ethernet |
Mechanical |
|
| Cabinet: | metal, protection standard IP 55 (NEMA 4X, others as an option) |
| Dimensions: | from 553 x 600 x 500 mm (WxHxD) to 800 x 2000 x 600 mm; according to channel number and optional equipment (air conditioning) |
| Weight: | approx. 35 - 175 kg; depending on number of channels and optional equipment |
Software |
|
| System requirement: | PC with Windows XP, Vista, 7 |
| Display of measuring data: | actual readings numerically, filed data as time-concentration graphs |
![]() |