
Mercury Monitoring system for Natural Gas
Automatic and Continuous Monitoring of Mercury Levels in Natural Gas
Benefits of the MMS-NG Mercury Monitoring System:
- Automatic and Continuous Operation
- Fast and Reliable Results
- Detects Elemental and Bound Mercury
- Automatic Calibration
- Sample Point Multiplexer
- Sample Dilution for High Concentrations
- No Carrier Gases Required
- Certified for Hazardous Zones
Preface of Mercury Monitor
Natural gas often contains mercury at concentrations that vary from below 1 to above 10000 µg/m³. Mercury is both toxic and potentially damaging: it can corrode or embrittle common gas plant components. Gas plants reduce mercury in natural gas with mercury removal units (MRUs). MRUs use fixed bed absorbers, often with sulfur-impregnated carbon or other chemisorbents as the active material. The Mercury Monitoring System is an ideal tool to determine the efficiency of each MRU in real time, and necessary to successfully monitor and control mercury concentrations during natural gas production and processing.
Continuous Measurement of Mercury in Natural Gas
Mercury Instruments has
engineered a system for
automatic and continuous monitoring of mercury levels in natural gas
and other flammable gases. We install the mercury analyzer (and
accessories such as a calibrator) in a pressurized enclosure that is
approved for use even in hazardous zones (ATEX, EExP). A heated
system takes the sample and guides it to the analyzer via stainless
steel tubing that has been surface-treated for ultra-low adsorptive.
A built-in gas sensor will shut the system down and stop the sample
gas flow if any leakage is detected.
Highly Selective and Sensitive Detector
The UT-3000 Mercury
Ultratracer is used to detect and precisely measure mercury
levels in natural gas. A proprietary gold collector selectively
absorbs mercury from a constant sample volume. The collector is then
heated for a very short period. The mercury is thus released into an
optical cell and quantitatively detected using atomic absorption
measurement (AAS). In contrast to atomic fluorescence systems, the
UT-3000 does not require special carrier gases. Air is used as a
carrier gas, thus keeping the gold surface clean and enhancing the
collector lifetime compared to other systems. With the UT-3000,
interference from hydrocarbons or other gaseous components is
insignificant (see Benzene interference diagram).
Benzene interference of different mercury detectors

Sampling System for Natural Gas
The integrity of the sampling system is as
important as the analyzer itself. To sample natural gas from a
pipeline, the sampling system must reduce pressure and guide the
sample from the sampling point to the analyzer, leaving the mercury
concentration unchanged. Plus it should show a minimum lag time and
be suited to hard use in hazardous zones. The MMS sample
conditioning system by Mercury Instruments fulfills these
requirements. The surface of the pressure-reduction system is
electrically heated to obviate condensation and mercury loss caused
by the Joule-Thomson cooling effect. A specially coated coalescing
filter effectively removes aqueous mist as well as hydrocarbon
condensate. Tubing and filter surfaces are specially coated for
ultra-low adsorption and constantly conditioned with
sample gas. Our system design allows a maximum input pressure of
3480 psig (240 bar) and the output pressure is adjustable from 1 to
28 psig (0.07 to 2 bar).
Sampling Point Multiplexer
For process control it is often useful to monitor mercury levels at different process points - the inlet and outlet of each mercury removal system, for example. Our microprocessor-controlled multiplexer unit feeds samples from up to 16 sampling points to the analyzer for sequential measurement. Using the fast loop principle all sample lines, including those currently not being measured, are continuously purged with sample and thereby constantly preconditioned. A short response time is the result.
Sample point multiplexer
unit. Up to 16 sample streams can be connected to the multiplexer
inlet. For minimum response times, the sample lines are continuously
purged to keep them conditioned. Specially coated bypass filters
retain entrained liquids (coalescing filters can also be used).

Example: the MMS Mercury Monitoring System installed in a gas separation plant, where mercury must be removed from the gas to avoid corrosion of aluminum heat exchangers and to produce mercury-free product streams. The MMS is the perfect tool to monitor the efficiency of the mercury removal units and to verify a low mercury concentration in the processed gas.
Calibration and Quality Assurance
For quality assurance
a regular calibration check is recommended. Calibration can
be performed manually using the Manual Calibration Set or
automatically with the Automatic Calibration Unit. Both devices use
the principle of static calibration with mercury-saturated air.
These devices are maintenance-free and unlike permeation devices do
not require re-weighing.<p>Calibration gas is
generated in a mercury chamber surrounded by a thermoelectrically
cooled aluminum jacket. The chamber contains highly pure elemental
mercury, and the temperature of the mercury is precisely measured by
a sensor. The UT-3000 analyzer uses this temperature signal to
calculate the true mercury
concentration via the mercury vapor pressure equation recommended by
the NIST*. A small volume of air is extracted from the mercury
chamber with a syringe and injected into the analyzer’s calibration
port. The Automatic Calibration Unit is permanently installed in the
analyzer cabinet, and it extracts the mercury vapor and injects it
into the UT-3000 calibration port by means of a digital syringe. Up
to 12 calibration points can be programmed, and automatic
calibration can be set to repeat at a fixed time
interval or at a fixed daily hour. * National Institute of
Standards and Technology
To the right you will see a manual calibration of the mercury analyzer. A known volume of calibration gas is drawn from the calibration device using a gas-tight micro-syringe and injected into the analyzer’s calibration port

MMS-NG Mercury Monitoring Specifications
Multiplexer Module
- Number of sampling points: 2-16
- Purging of sample lines: continuously
- Measuring duration for one measurement: approx. 3 minutes
Detector
- Detection principle: Atomic Absorption Spectrometry (AAS)
- Analytical wavelength: 253.7 nm
- Measuring range (standard): 0.1 to 100 µg/m³; 1 to 1000 µg/m³; 1 to 2000 µg/m³
- Matrix effect elimination: GoldTrap technique
- Measuring range: 0.001 µg/m³ … 50 µg/m³ (with sample dilution system up to 2000 µg/m³
- Signal Output: 4-20 mA; RS232; Modbus RTU RS485, Ethernet
Sample Conditioning System (SCS)
- Max. inlet pressure: 240 bar (3480 psi)
- Pressure regulator: heated (EEx certified)
- Sample wetted surfaces: coated for low mercury adsorption
- Weight: approx. 35 - 175 kg; depending on number of channels and optional equipment
Automatic Calibrator
- Operating principle: Mercury vapor saturation, injection of a constant volume
- Display of measuring data: actual readings numerically, filed data as time-concentration graphs
- Calculation of mercury vapor pressure: according to NIST recommended equation
Certification and Applied Standards
- Operating principle: Mercury vapor saturation, injection of a constant volume
- Display of measuring data: actual readings numerically, filed data as time-concentration graphs
- Calculation of mercury vapor pressure: according to NIST recommended equation
Certification and Applied Standards
- Hazardous zone certification: ATEX 2G IIC T4 EExp (or equivalent)
- Calibration: ISO/DIS 6978-3; ASTM D 5954; VDI 2267 Part 8; NIST recommendations
- Sampling and mercury determination: ISO 6978-1; ASTM 5954

